Physics of Capnography
Nitrous oxide absorbs IR light.
Absorption is minimal at IR 4.3 µ m
Collision Broadening Phenomenon
Effect of n2o on C02 measurements |
Use of 4.3 µ m IR light does not affect C02 measurements. |
Collision broadening phenomena does increase C02 values. |
Correction factors for the presence of nitrous oxide.
Percent Nitrous oxide |
Corrected reading = |
70 % |
Observed PC02 x 0.90 |
50% |
Observed PC02 x 0.94 |
References:
1. Raemer DB, Calalang I. Accuracy of end-tidal carbon dioxide tension analyzer. J Clin Monit 1991;7:19-208.
2. Kennell EM, Andrews RW, Wollman H. Correction factors for nitrous oxide in the infrared analysis of carbon dioxide. Anesthesiology 1973;39:441-3.
3. Carbon dioxide monitors. Health Devices 1986;15:255-85.
4. Paloheimo M, Valli M, Ahjopalo H. A guide to C02 monitoring. Finland:Datex Instrumentaitrium, 1988.